YMC BioPro HIC HT is a hydrophobic interaction chromatography (HIC) column with butyl (C4) stationary phase bonded to hydrophilic polymer resin. YMC BioPro HIC HT is suitable for analysis of biopharmaceuticals such as antibody-drug conjugates (ADC’s) and drug-to-antibody ratio (DAR).
YMC BioPro HIC HT is a hydrophobic interaction chromatography (HIC) column with butyl (C4) stationary phase bonded to hydrophilic polymer resin. YMC BioPro HIC HT is suitable for analysis of biopharmaceuticals such as antibody-drug conjugates (ADC’s) and drug-to-antibody ratio (DAR).
BioPro HIC HT has high pressure tolerance with excellent chromatographic resolution by optimized packing condition. This enables rapid analysis without losing the resolution at high pressure, which is too high for other commercially available HIC column.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (85/15) |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
DAR analysis by using BioPro HIC HT and competitor columns under two mobile phase conditions varying 2-propanol content was carried out. The competitor A column having high hydrophobicity did not elute peak #5 completely with the mobile phase containing less than 15% 2-propanol. The competitor B column eluted all peaks with the 10% 2-propanol mobile phase condition, but it showed poor separation between peaks #4 and #5. On the other hand, BioPro HIC HT eluted all peaks completely with good separation and another peak was partially separated on the peak #3. Additionally, it is noteworthy that the same DAR values were observed at both mobile phases.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (90/10) or (85/15) 0-100%B (0-15 min), 100%B (15-20 min), 0%B (20-35 min) |
Flow rate | 0.5 mL/min |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
BioPro HIC HT has high pressure tolerance with excellent chromatographic resolution by optimized packing condition. This enables rapid analysis without losing the resolution at high pressure, which is too high for other commercially available HIC column.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (85/15) |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
DAR analysis by using BioPro HIC HT and competitor columns under two mobile phase conditions varying 2-propanol content was carried out. The competitor A column having high hydrophobicity did not elute peak #5 completely with the mobile phase containing less than 15% 2-propanol. The competitor B column eluted all peaks with the 10% 2-propanol mobile phase condition, but it showed poor separation between peaks #4 and #5. On the other hand, BioPro HIC HT eluted all peaks completely with good separation and another peak was partially separated on the peak #3. Additionally, it is noteworthy that the same DAR values were observed at both mobile phases.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (90/10) or (85/15) 0-100%B (0-15 min), 100%B (15-20 min), 0%B (20-35 min) |
Flow rate | 0.5 mL/min |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
BioPro HIC HT has high pressure tolerance with excellent chromatographic resolution by optimized packing condition. This enables rapid analysis without losing the resolution at high pressure, which is too high for other commercially available HIC column.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (85/15) |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
DAR analysis by using BioPro HIC HT and competitor columns under two mobile phase conditions varying 2-propanol content was carried out. The competitor A column having high hydrophobicity did not elute peak #5 completely with the mobile phase containing less than 15% 2-propanol. The competitor B column eluted all peaks with the 10% 2-propanol mobile phase condition, but it showed poor separation between peaks #4 and #5. On the other hand, BioPro HIC HT eluted all peaks completely with good separation and another peak was partially separated on the peak #3. Additionally, it is noteworthy that the same DAR values were observed at both mobile phases.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (90/10) or (85/15) 0-100%B (0-15 min), 100%B (15-20 min), 0%B (20-35 min) |
Flow rate | 0.5 mL/min |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
BioPro HIC HT has high pressure tolerance with excellent chromatographic resolution by optimized packing condition. This enables rapid analysis without losing the resolution at high pressure, which is too high for other commercially available HIC column.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (85/15) |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
DAR analysis by using BioPro HIC HT and competitor columns under two mobile phase conditions varying 2-propanol content was carried out. The competitor A column having high hydrophobicity did not elute peak #5 completely with the mobile phase containing less than 15% 2-propanol. The competitor B column eluted all peaks with the 10% 2-propanol mobile phase condition, but it showed poor separation between peaks #4 and #5. On the other hand, BioPro HIC HT eluted all peaks completely with good separation and another peak was partially separated on the peak #3. Additionally, it is noteworthy that the same DAR values were observed at both mobile phases.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (90/10) or (85/15) 0-100%B (0-15 min), 100%B (15-20 min), 0%B (20-35 min) |
Flow rate | 0.5 mL/min |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
BioPro HIC HT has high pressure tolerance with excellent chromatographic resolution by optimized packing condition. This enables rapid analysis without losing the resolution at high pressure, which is too high for other commercially available HIC column.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (85/15) |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
DAR analysis by using BioPro HIC HT and competitor columns under two mobile phase conditions varying 2-propanol content was carried out. The competitor A column having high hydrophobicity did not elute peak #5 completely with the mobile phase containing less than 15% 2-propanol. The competitor B column eluted all peaks with the 10% 2-propanol mobile phase condition, but it showed poor separation between peaks #4 and #5. On the other hand, BioPro HIC HT eluted all peaks completely with good separation and another peak was partially separated on the peak #3. Additionally, it is noteworthy that the same DAR values were observed at both mobile phases.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (90/10) or (85/15) 0-100%B (0-15 min), 100%B (15-20 min), 0%B (20-35 min) |
Flow rate | 0.5 mL/min |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
BioPro HIC HT has high pressure tolerance with excellent chromatographic resolution by optimized packing condition. This enables rapid analysis without losing the resolution at high pressure, which is too high for other commercially available HIC column.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (85/15) |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
DAR analysis by using BioPro HIC HT and competitor columns under two mobile phase conditions varying 2-propanol content was carried out. The competitor A column having high hydrophobicity did not elute peak #5 completely with the mobile phase containing less than 15% 2-propanol. The competitor B column eluted all peaks with the 10% 2-propanol mobile phase condition, but it showed poor separation between peaks #4 and #5. On the other hand, BioPro HIC HT eluted all peaks completely with good separation and another peak was partially separated on the peak #3. Additionally, it is noteworthy that the same DAR values were observed at both mobile phases.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (90/10) or (85/15) 0-100%B (0-15 min), 100%B (15-20 min), 0%B (20-35 min) |
Flow rate | 0.5 mL/min |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
BioPro HIC HT has high pressure tolerance with excellent chromatographic resolution by optimized packing condition. This enables rapid analysis without losing the resolution at high pressure, which is too high for other commercially available HIC column.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (85/15) |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
DAR analysis by using BioPro HIC HT and competitor columns under two mobile phase conditions varying 2-propanol content was carried out. The competitor A column having high hydrophobicity did not elute peak #5 completely with the mobile phase containing less than 15% 2-propanol. The competitor B column eluted all peaks with the 10% 2-propanol mobile phase condition, but it showed poor separation between peaks #4 and #5. On the other hand, BioPro HIC HT eluted all peaks completely with good separation and another peak was partially separated on the peak #3. Additionally, it is noteworthy that the same DAR values were observed at both mobile phases.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (90/10) or (85/15) 0-100%B (0-15 min), 100%B (15-20 min), 0%B (20-35 min) |
Flow rate | 0.5 mL/min |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
BioPro HIC HT has high pressure tolerance with excellent chromatographic resolution by optimized packing condition. This enables rapid analysis without losing the resolution at high pressure, which is too high for other commercially available HIC column.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (85/15) |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
DAR analysis by using BioPro HIC HT and competitor columns under two mobile phase conditions varying 2-propanol content was carried out. The competitor A column having high hydrophobicity did not elute peak #5 completely with the mobile phase containing less than 15% 2-propanol. The competitor B column eluted all peaks with the 10% 2-propanol mobile phase condition, but it showed poor separation between peaks #4 and #5. On the other hand, BioPro HIC HT eluted all peaks completely with good separation and another peak was partially separated on the peak #3. Additionally, it is noteworthy that the same DAR values were observed at both mobile phases.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (90/10) or (85/15) 0-100%B (0-15 min), 100%B (15-20 min), 0%B (20-35 min) |
Flow rate | 0.5 mL/min |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
BioPro HIC HT has high pressure tolerance with excellent chromatographic resolution by optimized packing condition. This enables rapid analysis without losing the resolution at high pressure, which is too high for other commercially available HIC column.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (85/15) |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
DAR analysis by using BioPro HIC HT and competitor columns under two mobile phase conditions varying 2-propanol content was carried out. The competitor A column having high hydrophobicity did not elute peak #5 completely with the mobile phase containing less than 15% 2-propanol. The competitor B column eluted all peaks with the 10% 2-propanol mobile phase condition, but it showed poor separation between peaks #4 and #5. On the other hand, BioPro HIC HT eluted all peaks completely with good separation and another peak was partially separated on the peak #3. Additionally, it is noteworthy that the same DAR values were observed at both mobile phases.
Column | 100 X 4.6 mmI.D. |
Eluent | A) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0) containing 1.0 M (NH4)2SO4 B) 20 mM NaH2PO4-Na2HPO4Â (pH 7.0)/2-propanol (90/10) or (85/15) 0-100%B (0-15 min), 100%B (15-20 min), 0%B (20-35 min) |
Flow rate | 0.5 mL/min |
Temperature | 25°C |
Detection | UV at 280 nm |
Injection | 10 µL |
Sample | Brentuximab vedotin (2.5 mg/mL) |
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.
Particle size (µm) |
Column size Inner diameter X length (mm) |
Product number |
---|---|---|
2.3 | 4.6 X 33 | BHH00SQ3-H346PTH |
4.6 X 100 | BHH00SQ3-1046PTH |
Please contact us for product with other sizes.