YMC-Triart Bio C4

YMC-Triart Bio C4

This organic/inorganic hybrid silica based column with butyl functional group is a wide pore colume suitable for separation of proteins and oligonucleotides as well as separation of intact antibodies under high temperature. Its excellent durability makes analysis up to 90°C at pH 1-7, and 50°C at pH 7-10 possible. It is also compatible with LC-MS.

YMC-Triart Bio C4

This organic/inorganic hybrid silica based column with butyl functional group is a wide pore colume suitable for separation of proteins and oligonucleotides as well as separation of intact antibodies under high temperature. Its excellent durability makes analysis up to 90°C at pH 1-7, and 50°C at pH 7-10 possible. It is also compatible with LC-MS.

Specification:

Base
Organic/inorganic hybrid silica
Base
Organic/inorganic hybrid silica
Stationary phase
USP L26, C4
Stationary phase
USP L26, C4
Particle size
1.9 µm, 3 µm, 5 µm
Particle size
1.9 µm, 3 µm, 5 µm
Pore size
300 Ã…
Pore size
300 Ã…
Usable pH range
1 ~10
Usable pH range
1 ~10
Endcapping
Yes
Endcapping
Yes

Specification:

Base
Organic/inorganic hybrid silica
Base
Organic/inorganic hybrid silica
Stationary phase
USP L26, C4
Stationary phase
USP L26, C4
Particle size
1.9 µm, 3 µm, 5 µm
Particle size
1.9 µm, 3 µm, 5 µm
Pore size
300 Ã…
Pore size
300 Ã…
Usable pH range
1 ~10
Usable pH range
1 ~10
Endcapping
Yes
Endcapping
Yes

Specification:

Base
Organic/inorganic hybrid silica
Base
Organic/inorganic hybrid silica
Stationary phase
USP L26, C4
Stationary phase
USP L26, C4
Particle size
1.9 µm, 3 µm, 5 µm
Particle size
1.9 µm, 3 µm, 5 µm
Pore size
300 Ã…
Pore size
300 Ã…
Usable pH range
1 ~10
Usable pH range
1 ~10
Endcapping
Yes
Endcapping
Yes

Useful for separation of intact antibodies under high temperature

In the analysis of intact antibodies from 40°C up to 90°C, good peak shape was obtained at 70°C and onwards. Its excellent durability enables analysis of antibodies even at high temperature up to 90°C.

Column YMC-Triart Bio C4 (3 µm, 300 Å)
150 X 3.0 mmI.D.
Eluent A) water/TFA (100/0.1)
B) acetonitrile/TFA (100/0.1)
30-60%B (0-15 min), 90%B (15-30 min)
Flow rate 0.4 mL/min
Detection UV at 220 nm
Sample Humanized monoclonal IgG1
Injection 4 µL

 

Good peak shape with mobile phase suitable for LC-MS

Good peak shape was obtained using YMC-Triart Bio C4 compared to competitor C4. As YMC-Triart Bio C4 is compatible with mobile phase containing formic acid, it is suitable for LC-MS analysis.

  1. Cytochrome c (Horse heart)
  2. Insulin (Bovine pancreas)
  3. Transferrin (Human)
  4. BSA
  5. B-Lactogloblin A (Bovine)
  6. a-Chymotrypsinogen A (Bovine pancreas)
  7. Ovalbumin
Column size 150 X 3.0 mmI.D or 4.6 mml.D.
Eluent  A) water/formic acid (100/0.1)
B) acetonitrile/formid acid (100/0.1)
10-95%B (0-15 min)
Flow rate 0.4 mL/min for 3.0 mml.D.
1.0 mL/min for 4.6 mml.D.
Temperature 40℃
Detection UV at 220 nm

Useful for separation of intact antibodies under high temperature

In the analysis of intact antibodies from 40°C up to 90°C, good peak shape was obtained at 70°C and onwards. Its excellent durability enables analysis of antibodies even at high temperature up to 90°C.

Column YMC-Triart Bio C4 (3 µm, 300 Å)
150 X 3.0 mmI.D.
Eluent A) water/TFA (100/0.1)
B) acetonitrile/TFA (100/0.1)
30-60%B (0-15 min), 90%B (15-30 min)
Flow rate 0.4 mL/min
Detection UV at 220 nm
Sample Humanized monoclonal IgG1
Injection 4 µL

 

Good peak shape with mobile phase suitable for LC-MS

Good peak shape was obtained using YMC-Triart Bio C4 compared to competitor C4. As YMC-Triart Bio C4 is compatible with mobile phase containing formic acid, it is suitable for LC-MS analysis.

  1. Cytochrome c (Horse heart)
  2. Insulin (Bovine pancreas)
  3. Transferrin (Human)
  4. BSA
  5. B-Lactogloblin A (Bovine)
  6. a-Chymotrypsinogen A (Bovine pancreas)
  7. Ovalbumin
Column size 150 X 3.0 mmI.D or 4.6 mml.D.
Eluent  A) water/formic acid (100/0.1)
B) acetonitrile/formid acid (100/0.1)
10-95%B (0-15 min)
Flow rate 0.4 mL/min for 3.0 mml.D.
1.0 mL/min for 4.6 mml.D.
Temperature 40℃
Detection UV at 220 nm

Useful for separation of intact antibodies under high temperature

In the analysis of intact antibodies from 40°C up to 90°C, good peak shape was obtained at 70°C and onwards. Its excellent durability enables analysis of antibodies even at high temperature up to 90°C.

Column YMC-Triart Bio C4 (3 µm, 300 Å)
150 X 3.0 mmI.D.
Eluent A) water/TFA (100/0.1)
B) acetonitrile/TFA (100/0.1)
30-60%B (0-15 min), 90%B (15-30 min)
Flow rate 0.4 mL/min
Detection UV at 220 nm
Sample Humanized monoclonal IgG1
Injection 4 µL

 

Good peak shape with mobile phase suitable for LC-MS

Good peak shape was obtained using YMC-Triart Bio C4 compared to competitor C4. As YMC-Triart Bio C4 is compatible with mobile phase containing formic acid, it is suitable for LC-MS analysis.

  1. Cytochrome c (Horse heart)
  2. Insulin (Bovine pancreas)
  3. Transferrin (Human)
  4. BSA
  5. B-Lactogloblin A (Bovine)
  6. a-Chymotrypsinogen A (Bovine pancreas)
  7. Ovalbumin
Column size 150 X 3.0 mmI.D or 4.6 mml.D.
Eluent  A) water/formic acid (100/0.1)
B) acetonitrile/formid acid (100/0.1)
10-95%B (0-15 min)
Flow rate 0.4 mL/min for 3.0 mml.D.
1.0 mL/min for 4.6 mml.D.
Temperature 40℃
Detection UV at 220 nm

Specification:

Base
Organic/inorganic hybrid silica
Base
Organic/inorganic hybrid silica
Stationary phase
USP L26, C4
Stationary phase
USP L26, C4
Particle size
1.9 µm, 3 µm, 5 µm
Particle size
1.9 µm, 3 µm, 5 µm
Pore size
300 Ã…
Pore size
300 Ã…
Usable pH range
1 ~10
Usable pH range
1 ~10
Endcapping
Yes
Endcapping
Yes

Specification:

Base
Organic/inorganic hybrid silica
Base
Organic/inorganic hybrid silica
Stationary phase
USP L26, C4
Stationary phase
USP L26, C4
Particle size
1.9 µm, 3 µm, 5 µm
Particle size
1.9 µm, 3 µm, 5 µm
Pore size
300 Ã…
Pore size
300 Ã…
Usable pH range
1 ~10
Usable pH range
1 ~10
Endcapping
Yes
Endcapping
Yes

Specification:

Base
Organic/inorganic hybrid silica
Base
Organic/inorganic hybrid silica
Stationary phase
USP L26, C4
Stationary phase
USP L26, C4
Particle size
1.9 µm, 3 µm, 5 µm
Particle size
1.9 µm, 3 µm, 5 µm
Pore size
300 Ã…
Pore size
300 Ã…
Usable pH range
1 ~10
Usable pH range
1 ~10
Endcapping
Yes
Endcapping
Yes

Useful for separation of intact antibodies under high temperature

In the analysis of intact antibodies from 40°C up to 90°C, good peak shape was obtained at 70°C and onwards. Its excellent durability enables analysis of antibodies even at high temperature up to 90°C.

Column YMC-Triart Bio C4 (3 µm, 300 Å)
150 X 3.0 mmI.D.
Eluent A) water/TFA (100/0.1)
B) acetonitrile/TFA (100/0.1)
30-60%B (0-15 min), 90%B (15-30 min)
Flow rate 0.4 mL/min
Detection UV at 220 nm
Sample Humanized monoclonal IgG1
Injection 4 µL

 

Good peak shape with mobile phase suitable for LC-MS

Good peak shape was obtained using YMC-Triart Bio C4 compared to competitor C4. As YMC-Triart Bio C4 is compatible with mobile phase containing formic acid, it is suitable for LC-MS analysis.

  1. Cytochrome c (Horse heart)
  2. Insulin (Bovine pancreas)
  3. Transferrin (Human)
  4. BSA
  5. B-Lactogloblin A (Bovine)
  6. a-Chymotrypsinogen A (Bovine pancreas)
  7. Ovalbumin
Column size 150 X 3.0 mmI.D or 4.6 mml.D.
Eluent  A) water/formic acid (100/0.1)
B) acetonitrile/formid acid (100/0.1)
10-95%B (0-15 min)
Flow rate 0.4 mL/min for 3.0 mml.D.
1.0 mL/min for 4.6 mml.D.
Temperature 40℃
Detection UV at 220 nm

Useful for separation of intact antibodies under high temperature

In the analysis of intact antibodies from 40°C up to 90°C, good peak shape was obtained at 70°C and onwards. Its excellent durability enables analysis of antibodies even at high temperature up to 90°C.

Column YMC-Triart Bio C4 (3 µm, 300 Å)
150 X 3.0 mmI.D.
Eluent A) water/TFA (100/0.1)
B) acetonitrile/TFA (100/0.1)
30-60%B (0-15 min), 90%B (15-30 min)
Flow rate 0.4 mL/min
Detection UV at 220 nm
Sample Humanized monoclonal IgG1
Injection 4 µL

 

Good peak shape with mobile phase suitable for LC-MS

Good peak shape was obtained using YMC-Triart Bio C4 compared to competitor C4. As YMC-Triart Bio C4 is compatible with mobile phase containing formic acid, it is suitable for LC-MS analysis.

  1. Cytochrome c (Horse heart)
  2. Insulin (Bovine pancreas)
  3. Transferrin (Human)
  4. BSA
  5. B-Lactogloblin A (Bovine)
  6. a-Chymotrypsinogen A (Bovine pancreas)
  7. Ovalbumin
Column size 150 X 3.0 mmI.D or 4.6 mml.D.
Eluent  A) water/formic acid (100/0.1)
B) acetonitrile/formid acid (100/0.1)
10-95%B (0-15 min)
Flow rate 0.4 mL/min for 3.0 mml.D.
1.0 mL/min for 4.6 mml.D.
Temperature 40℃
Detection UV at 220 nm

Useful for separation of intact antibodies under high temperature

In the analysis of intact antibodies from 40°C up to 90°C, good peak shape was obtained at 70°C and onwards. Its excellent durability enables analysis of antibodies even at high temperature up to 90°C.

Column YMC-Triart Bio C4 (3 µm, 300 Å)
150 X 3.0 mmI.D.
Eluent A) water/TFA (100/0.1)
B) acetonitrile/TFA (100/0.1)
30-60%B (0-15 min), 90%B (15-30 min)
Flow rate 0.4 mL/min
Detection UV at 220 nm
Sample Humanized monoclonal IgG1
Injection 4 µL

 

Good peak shape with mobile phase suitable for LC-MS

Good peak shape was obtained using YMC-Triart Bio C4 compared to competitor C4. As YMC-Triart Bio C4 is compatible with mobile phase containing formic acid, it is suitable for LC-MS analysis.

  1. Cytochrome c (Horse heart)
  2. Insulin (Bovine pancreas)
  3. Transferrin (Human)
  4. BSA
  5. B-Lactogloblin A (Bovine)
  6. a-Chymotrypsinogen A (Bovine pancreas)
  7. Ovalbumin
Column size 150 X 3.0 mmI.D or 4.6 mml.D.
Eluent  A) water/formic acid (100/0.1)
B) acetonitrile/formid acid (100/0.1)
10-95%B (0-15 min)
Flow rate 0.4 mL/min for 3.0 mml.D.
1.0 mL/min for 4.6 mml.D.
Temperature 40℃
Detection UV at 220 nm

Specification:

Base
Organic/inorganic hybrid silica
Base
Organic/inorganic hybrid silica
Stationary phase
USP L26, C4
Stationary phase
USP L26, C4
Particle size
1.9 µm, 3 µm, 5 µm
Particle size
1.9 µm, 3 µm, 5 µm
Pore size
300 Ã…
Pore size
300 Ã…
Usable pH range
1 ~10
Usable pH range
1 ~10
Endcapping
Yes
Endcapping
Yes

Specification:

Base
Organic/inorganic hybrid silica
Base
Organic/inorganic hybrid silica
Stationary phase
USP L26, C4
Stationary phase
USP L26, C4
Particle size
1.9 µm, 3 µm, 5 µm
Particle size
1.9 µm, 3 µm, 5 µm
Pore size
300 Ã…
Pore size
300 Ã…
Usable pH range
1 ~10
Usable pH range
1 ~10
Endcapping
Yes
Endcapping
Yes

Specification:

Base
Organic/inorganic hybrid silica
Base
Organic/inorganic hybrid silica
Stationary phase
USP L26, C4
Stationary phase
USP L26, C4
Particle size
1.9 µm, 3 µm, 5 µm
Particle size
1.9 µm, 3 µm, 5 µm
Pore size
300 Ã…
Pore size
300 Ã…
Usable pH range
1 ~10
Usable pH range
1 ~10
Endcapping
Yes
Endcapping
Yes

Useful for separation of intact antibodies under high temperature

In the analysis of intact antibodies from 40°C up to 90°C, good peak shape was obtained at 70°C and onwards. Its excellent durability enables analysis of antibodies even at high temperature up to 90°C.

Column YMC-Triart Bio C4 (3 µm, 300 Å)
150 X 3.0 mmI.D.
Eluent A) water/TFA (100/0.1)
B) acetonitrile/TFA (100/0.1)
30-60%B (0-15 min), 90%B (15-30 min)
Flow rate 0.4 mL/min
Detection UV at 220 nm
Sample Humanized monoclonal IgG1
Injection 4 µL

 

Good peak shape with mobile phase suitable for LC-MS

Good peak shape was obtained using YMC-Triart Bio C4 compared to competitor C4. As YMC-Triart Bio C4 is compatible with mobile phase containing formic acid, it is suitable for LC-MS analysis.

  1. Cytochrome c (Horse heart)
  2. Insulin (Bovine pancreas)
  3. Transferrin (Human)
  4. BSA
  5. B-Lactogloblin A (Bovine)
  6. a-Chymotrypsinogen A (Bovine pancreas)
  7. Ovalbumin
Column size 150 X 3.0 mmI.D or 4.6 mml.D.
Eluent  A) water/formic acid (100/0.1)
B) acetonitrile/formid acid (100/0.1)
10-95%B (0-15 min)
Flow rate 0.4 mL/min for 3.0 mml.D.
1.0 mL/min for 4.6 mml.D.
Temperature 40℃
Detection UV at 220 nm

Useful for separation of intact antibodies under high temperature

In the analysis of intact antibodies from 40°C up to 90°C, good peak shape was obtained at 70°C and onwards. Its excellent durability enables analysis of antibodies even at high temperature up to 90°C.

Column YMC-Triart Bio C4 (3 µm, 300 Å)
150 X 3.0 mmI.D.
Eluent A) water/TFA (100/0.1)
B) acetonitrile/TFA (100/0.1)
30-60%B (0-15 min), 90%B (15-30 min)
Flow rate 0.4 mL/min
Detection UV at 220 nm
Sample Humanized monoclonal IgG1
Injection 4 µL

 

Good peak shape with mobile phase suitable for LC-MS

Good peak shape was obtained using YMC-Triart Bio C4 compared to competitor C4. As YMC-Triart Bio C4 is compatible with mobile phase containing formic acid, it is suitable for LC-MS analysis.

  1. Cytochrome c (Horse heart)
  2. Insulin (Bovine pancreas)
  3. Transferrin (Human)
  4. BSA
  5. B-Lactogloblin A (Bovine)
  6. a-Chymotrypsinogen A (Bovine pancreas)
  7. Ovalbumin
Column size 150 X 3.0 mmI.D or 4.6 mml.D.
Eluent  A) water/formic acid (100/0.1)
B) acetonitrile/formid acid (100/0.1)
10-95%B (0-15 min)
Flow rate 0.4 mL/min for 3.0 mml.D.
1.0 mL/min for 4.6 mml.D.
Temperature 40℃
Detection UV at 220 nm

Useful for separation of intact antibodies under high temperature

In the analysis of intact antibodies from 40°C up to 90°C, good peak shape was obtained at 70°C and onwards. Its excellent durability enables analysis of antibodies even at high temperature up to 90°C.

Column YMC-Triart Bio C4 (3 µm, 300 Å)
150 X 3.0 mmI.D.
Eluent A) water/TFA (100/0.1)
B) acetonitrile/TFA (100/0.1)
30-60%B (0-15 min), 90%B (15-30 min)
Flow rate 0.4 mL/min
Detection UV at 220 nm
Sample Humanized monoclonal IgG1
Injection 4 µL

 

Good peak shape with mobile phase suitable for LC-MS

Good peak shape was obtained using YMC-Triart Bio C4 compared to competitor C4. As YMC-Triart Bio C4 is compatible with mobile phase containing formic acid, it is suitable for LC-MS analysis.

  1. Cytochrome c (Horse heart)
  2. Insulin (Bovine pancreas)
  3. Transferrin (Human)
  4. BSA
  5. B-Lactogloblin A (Bovine)
  6. a-Chymotrypsinogen A (Bovine pancreas)
  7. Ovalbumin
Column size 150 X 3.0 mmI.D or 4.6 mml.D.
Eluent  A) water/formic acid (100/0.1)
B) acetonitrile/formid acid (100/0.1)
10-95%B (0-15 min)
Flow rate 0.4 mL/min for 3.0 mml.D.
1.0 mL/min for 4.6 mml.D.
Temperature 40℃
Detection UV at 220 nm

Related Product

Pro Series

YMC-Pack Pro C18

Pro Series

YMC-Pack Pro C18

Pro Series

YMC-Pack Pro C18

Packing Material

CHIRAL ART Coated type

Packing Material

CHIRAL ART Coated type

Packing Material

CHIRAL ART Coated type

Packing Material

CHIRAL ART Immobilized type

Packing Material

CHIRAL ART Immobilized type

Packing Material

CHIRAL ART Immobilized type

YMC-Triart

Triart Prep SIL

YMC-Triart

Triart Prep SIL

YMC-Triart

Triart Prep SIL

YMC-Triart

Triart Prep Bio200 C8

YMC-Triart

Triart Prep Bio200 C8

YMC-Triart

Triart Prep Bio200 C8

YMC-Triart

Triart Prep C8-S

YMC-Triart

Triart Prep C8-S

YMC-Triart

Triart Prep C8-S