YMC-Triart [Metal free column]

YMC-Triart [Metal free column]

The YMC-Triart stationary phases are packed in PEEK-lined stainless steel column hardware, complete with PEEK frits. It is designed to eliminate peak shape deterioration caused by metal components. It is suitable for analysis of coordination compounds. High pressure tolerance is guaranteed.

https://www.ymc.co.jp/data/download/Triart_RP_E.pdf
https://www.ymc.co.jp/data/download/Triart_Diol-HILIC_E.pdf
https://www.ymc.co.jp/data/download/metal_free_column_connection_e.pdf

YMC-Triart [Metal free column]

The YMC-Triart stationary phases are packed in PEEK-lined stainless steel column hardware, complete with PEEK frits. It is designed to eliminate peak shape deterioration caused by metal components. It is suitable for analysis of coordination compounds. High pressure tolerance is guaranteed.

Specification:

Wetted part (liner)
PEEK
Wetted part (liner)
PEEK
Outer tube
Stainless steel
Outer tube
Stainless steel
Frit
PEEK
Frit
PEEK

Specification:

Wetted part (liner)
PEEK
Wetted part (liner)
PEEK
Outer tube
Stainless steel
Outer tube
Stainless steel
Frit
PEEK
Frit
PEEK

Specification:

Wetted part (liner)
PEEK
Wetted part (liner)
PEEK
Outer tube
Stainless steel
Outer tube
Stainless steel
Frit
PEEK
Frit
PEEK

Improved Sensitivity on Coordination Compounds Analysis

Metal coordination compounds which have phosphate group in their structure tend to show poor peak shape by interacting with metal, such as stainless steel column. In this nucleotides analysis, better peak shape was achieved with the metal free column. The metal free column is also effective for highly sensitive analysis using LC-MS.

 

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

 

Elimination of sample-metal interaction by optimization of system configuration

Tubing or HPLC system components could induce peak shape deterioration. Nucleotides were analyzed on an ordinary HPLC system and a “non-metal” HPLC system which consists of a PEEK sample loop, a PEEK injector needle seat and a PEEK tubing between injector and column. The peak shape of coordination compounds can be greatly improved with these PEEK parts.

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

Improved Sensitivity on Coordination Compounds Analysis

Metal coordination compounds which have phosphate group in their structure tend to show poor peak shape by interacting with metal, such as stainless steel column. In this nucleotides analysis, better peak shape was achieved with the metal free column. The metal free column is also effective for highly sensitive analysis using LC-MS.

 

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

 

Elimination of sample-metal interaction by optimization of system configuration

Tubing or HPLC system components could induce peak shape deterioration. Nucleotides were analyzed on an ordinary HPLC system and a “non-metal” HPLC system which consists of a PEEK sample loop, a PEEK injector needle seat and a PEEK tubing between injector and column. The peak shape of coordination compounds can be greatly improved with these PEEK parts.

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

Improved Sensitivity on Coordination Compounds Analysis

Metal coordination compounds which have phosphate group in their structure tend to show poor peak shape by interacting with metal, such as stainless steel column. In this nucleotides analysis, better peak shape was achieved with the metal free column. The metal free column is also effective for highly sensitive analysis using LC-MS.

 

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

 

Elimination of sample-metal interaction by optimization of system configuration

Tubing or HPLC system components could induce peak shape deterioration. Nucleotides were analyzed on an ordinary HPLC system and a “non-metal” HPLC system which consists of a PEEK sample loop, a PEEK injector needle seat and a PEEK tubing between injector and column. The peak shape of coordination compounds can be greatly improved with these PEEK parts.

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

Specification:

Wetted part (liner)
PEEK
Wetted part (liner)
PEEK
Outer tube
Stainless steel
Outer tube
Stainless steel
Frit
PEEK
Frit
PEEK

Specification:

Wetted part (liner)
PEEK
Wetted part (liner)
PEEK
Outer tube
Stainless steel
Outer tube
Stainless steel
Frit
PEEK
Frit
PEEK

Specification:

Wetted part (liner)
PEEK
Wetted part (liner)
PEEK
Outer tube
Stainless steel
Outer tube
Stainless steel
Frit
PEEK
Frit
PEEK

Improved Sensitivity on Coordination Compounds Analysis

Metal coordination compounds which have phosphate group in their structure tend to show poor peak shape by interacting with metal, such as stainless steel column. In this nucleotides analysis, better peak shape was achieved with the metal free column. The metal free column is also effective for highly sensitive analysis using LC-MS.

 

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

 

Elimination of sample-metal interaction by optimization of system configuration

Tubing or HPLC system components could induce peak shape deterioration. Nucleotides were analyzed on an ordinary HPLC system and a “non-metal” HPLC system which consists of a PEEK sample loop, a PEEK injector needle seat and a PEEK tubing between injector and column. The peak shape of coordination compounds can be greatly improved with these PEEK parts.

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

Improved Sensitivity on Coordination Compounds Analysis

Metal coordination compounds which have phosphate group in their structure tend to show poor peak shape by interacting with metal, such as stainless steel column. In this nucleotides analysis, better peak shape was achieved with the metal free column. The metal free column is also effective for highly sensitive analysis using LC-MS.

 

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

 

Elimination of sample-metal interaction by optimization of system configuration

Tubing or HPLC system components could induce peak shape deterioration. Nucleotides were analyzed on an ordinary HPLC system and a “non-metal” HPLC system which consists of a PEEK sample loop, a PEEK injector needle seat and a PEEK tubing between injector and column. The peak shape of coordination compounds can be greatly improved with these PEEK parts.

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

Improved Sensitivity on Coordination Compounds Analysis

Metal coordination compounds which have phosphate group in their structure tend to show poor peak shape by interacting with metal, such as stainless steel column. In this nucleotides analysis, better peak shape was achieved with the metal free column. The metal free column is also effective for highly sensitive analysis using LC-MS.

 

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

 

Elimination of sample-metal interaction by optimization of system configuration

Tubing or HPLC system components could induce peak shape deterioration. Nucleotides were analyzed on an ordinary HPLC system and a “non-metal” HPLC system which consists of a PEEK sample loop, a PEEK injector needle seat and a PEEK tubing between injector and column. The peak shape of coordination compounds can be greatly improved with these PEEK parts.

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

Specification:

Wetted part (liner)
PEEK
Wetted part (liner)
PEEK
Outer tube
Stainless steel
Outer tube
Stainless steel
Frit
PEEK
Frit
PEEK

Specification:

Wetted part (liner)
PEEK
Wetted part (liner)
PEEK
Outer tube
Stainless steel
Outer tube
Stainless steel
Frit
PEEK
Frit
PEEK

Specification:

Wetted part (liner)
PEEK
Wetted part (liner)
PEEK
Outer tube
Stainless steel
Outer tube
Stainless steel
Frit
PEEK
Frit
PEEK

Improved Sensitivity on Coordination Compounds Analysis

Metal coordination compounds which have phosphate group in their structure tend to show poor peak shape by interacting with metal, such as stainless steel column. In this nucleotides analysis, better peak shape was achieved with the metal free column. The metal free column is also effective for highly sensitive analysis using LC-MS.

 

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

 

Elimination of sample-metal interaction by optimization of system configuration

Tubing or HPLC system components could induce peak shape deterioration. Nucleotides were analyzed on an ordinary HPLC system and a “non-metal” HPLC system which consists of a PEEK sample loop, a PEEK injector needle seat and a PEEK tubing between injector and column. The peak shape of coordination compounds can be greatly improved with these PEEK parts.

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

Improved Sensitivity on Coordination Compounds Analysis

Metal coordination compounds which have phosphate group in their structure tend to show poor peak shape by interacting with metal, such as stainless steel column. In this nucleotides analysis, better peak shape was achieved with the metal free column. The metal free column is also effective for highly sensitive analysis using LC-MS.

 

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

 

Elimination of sample-metal interaction by optimization of system configuration

Tubing or HPLC system components could induce peak shape deterioration. Nucleotides were analyzed on an ordinary HPLC system and a “non-metal” HPLC system which consists of a PEEK sample loop, a PEEK injector needle seat and a PEEK tubing between injector and column. The peak shape of coordination compounds can be greatly improved with these PEEK parts.

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

Improved Sensitivity on Coordination Compounds Analysis

Metal coordination compounds which have phosphate group in their structure tend to show poor peak shape by interacting with metal, such as stainless steel column. In this nucleotides analysis, better peak shape was achieved with the metal free column. The metal free column is also effective for highly sensitive analysis using LC-MS.

 

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

 

Elimination of sample-metal interaction by optimization of system configuration

Tubing or HPLC system components could induce peak shape deterioration. Nucleotides were analyzed on an ordinary HPLC system and a “non-metal” HPLC system which consists of a PEEK sample loop, a PEEK injector needle seat and a PEEK tubing between injector and column. The peak shape of coordination compounds can be greatly improved with these PEEK parts.

Column YMC-Triart C18, 3 μm, 50 x 2.1mmI.D.
Eluent 5 mM HCOONH4
Flow rate 0.21 mL/min
Temperature 25°C
Detection UV at 265 nm
Injection 1 μL (10 μg/mL)

Related Product

Packing Material

CHIRAL ART Coated type

Packing Material

CHIRAL ART Coated type

Packing Material

CHIRAL ART Coated type

Packing Material

CHIRAL ART Immobilized type

Packing Material

CHIRAL ART Immobilized type

Packing Material

CHIRAL ART Immobilized type

Columns

YMC-Pack PVA-Sil

Columns

YMC-Pack PVA-Sil

Columns

YMC-Pack PVA-Sil

Columns

YMC-Pack PA-G

Columns

YMC-Pack PA-G

Columns

YMC-Pack PA-G

Columns

YMC-Pack NH2

Columns

YMC-Pack NH2

Columns

YMC-Pack NH2

Columns

YMC-Pack Polyamine II

Columns

YMC-Pack Polyamine II

Columns

YMC-Pack Polyamine II